Pt100 / Pt1000 RTD Calculator

Convert platinum RTD resistance to temperature, or temperature to the resistance you should measure, using the standard IEC 60751 curve (α = 0.00385).

IEC 60751 platinum curve, α = 0.00385. Readings measured with a 2-wire connection include lead resistance; subtract it first.

Result

Temperature100 °C
Temperature212 °F
Resistance138.505 Ω
Sensitivity near this point0.3793 Ω/°C

The Pt100 curve

A Pt100 is a platinum resistance element with 100 Ω at 0 °C (a Pt1000 has 1000 Ω). Its resistance rises by roughly 0.385 Ω per °C, but not perfectly linearly, so the Callendar–Van Dusen equation is used:

R(T) = R₀ × (1 + A·T + B·T²)   for 0 to 850 °C
R(T) = R₀ × (1 + A·T + B·T² + C·(T − 100)·T³)   for −200 to 0 °C
A = 3.9083 × 10⁻³, B = −5.775 × 10⁻⁷, C = −4.183 × 10⁻¹²

Quick reference (Pt100)

Temperature (°C)Resistance (Ω)
-20018.52
-10060.26
-5080.31
0100.00
25109.73
50119.40
100138.51
150157.33
200175.86
300212.05
400247.09
500280.98

2-, 3- and 4-wire connections

The transmitter measures the element and the lead wires. With a 2-wire connection, lead resistance adds directly to the reading: 1 Ω of total lead resistance on a Pt100 reads about 2.6 °C high. A 3-wire connection compensates for the leads as long as all three wires have equal resistance, and a 4-wire connection eliminates lead resistance from the measurement entirely.

Field check

To check an RTD, disconnect it and measure its resistance with an ohmmeter (subtracting the lead resistance on 2-wire runs), then compare with the calculator at the known process or bath temperature. An open element reads infinite resistance; a shorted element reads near 0 Ω, which a transmitter will interpret as far below −200 °C.